Power Turn Elite™ Lateral Rotation Mattress System- Model 9800

Lateral Rotation Therapy with True Low Air Loss

The Power-Turn™ Elite Mattress System combines two clinically effective therapies (three clinically effective therapies when pulsation mode is added): Lateral Rotation and True Low Air Loss in a portable system. The mattress utilizes 18 independent 9″ deep air cells to provide proper pressure redistribution for the prevention and treatment of Stages I-IV pressure ulcers. Lateral Rotation Therapy reduces the risk of pulmonary complications for patients who are immobile. A Full, Left, or Right Turn mode is offered with either a 25º Small Turn or a 40º Big Turn. The mattress replacement is designed to fit all standard hospital beds and includes a high-density foam base, head bolster, and abduction wedge for additional patient safety. The Power-Turn™ Elite Lateral Rotation Mattress System supports patients weighing from 650 lbs. to 1000 lbs. The Power-Turn™ Elite meets California Technical Bulletin #117 for fire retardancy. An optional fire barrier is available to meet the most stringent fire safety requirements when requested for foam base. Lateral Rotation and True Low Air Loss. The Power-Turn® Elite Model 9800

Bladders:
Ventilated low air loss cells have air holes throughout the mattress to help minimize skin maceration by reducing excess moisture under the patient. Air cells are constructed of TPU, thermal polyurethane material. This combination ensures strength, durability, and patient comfort.

Cover:
Vyvex – III™ – Urethane coated, multi-stretch, moisture-resistant / vapor-permeable, low shear, quilted cover, transfers moisture away from the patient’s skin. The cover is washable, anti-microbial, fluid-resistant, and fire-retardant.

Hose connections are color-coded for easy connection.
The Power-Turn® Elite Pump Model 9801 Blower Pump:
The blower pump is controlled by a micro-processor which allows the caregiver total control of comfort and chosen therapy. Cycle time ranges from 5 to 95 minutes. The blower pump inflates the mattress within minutes. This reduces the waiting time for therapeutic treatment to begin.Blower Options:
– Pulsating mode allows caregivers to provide pulsating low air loss therapy.
– Automatic 2 position fowler sensor detects patient positioning and automatically increases power as needed in the sacral area. 10% adjustment for a 30º incline. 5% additional adjustment for a 40º incline.
The Power-Turn® Elite Head Rest and Leg Abductor Head Rest and Leg Abductor:
The headrest and leg abductor wedge provide support during the turning cycles. Head support is constructed from Viscotec™, a viscoelastic foam. The leg abductor is constructed from PPF™, Pressurized Polymer Foam. Both are covered with Vyvex-III™.
MODEL 9800
Power-Turn Elite ™ – Lateral Rotation Therapy Mattress and Pump
Dimensions:
36″ x 80″ x 9″
Weight:
23 lbs.
Air Cells:
18 – 9 ” deep cells
Ventilated low air loss bladders with air holes throughout the mattress.
Side Bolster :
Patented air side rail bolsters are built into the mattress perimeter which supports the patient and ensures patient safety during rotation cycles.
Mattress Cover:
Vyvex-III™ – Urethane coated, multi-stretch, low shear, moisture vapor permeable, quilted, removable, zippered cover, is anti-microbial, and meets Cal. Tech. #117 for fire-retardancy.
Degree of Turn:
25º Small turn
40º Big turn
WARRANTY
Mattress:
12 month limited
Pump:
12 month limited
Cover:
6 month limited
ORDERING INFORMATION
Model 9800
36″ x 80″ x 9″ , 650 lbs capacity
Model 9842
42″ x 80″ x 9″ , 800 lbs capacity
Model 9848
48″ x 80″ x 9″ , 1000 lbs capacity
Model 9800-AW
Abductor Wedge
Model 9800-HR
Head Rest
Model 9826
Replacement zippered cover
Model 9826-42
42″ Replacement zippered cover
Model 9826-48
48″ Replacement zippered cover
Custom and bariatric sizes are available. Please state the patient’s weight when ordering
Use of side rails is recommended for patient safety.
PDAC APPROVED
Medicare Code E0277
Power Turn™ Elite meets Group 2 Support Surface requirements for Medicare EO277. Patient must meet all medical criteria to qualify for reimbursement.
MODEL 9801
Power-Turn Elite ™ Lateral Rotation Therapy- Pump only
Dimensions:
17″L x 7″W x 10.5″H
Weight:
12 lbs.
Electrical:
120V 60 hz
Listings:
UL® 544 Medical Equipment
Cycle Time:
5 – 95 Minutes
Air Flow:
Over 1800 Liters per minute
Current:
5A
Ground Resistance:
0.1 ohms maximum
Power Cord:
12 ft. cord with hospital grade plug
Modes of Operation :
Auto full turn
Left turn
Right turn
Auto-firm
Static

New Products

Tradewind ATS-RC with Remote ControlAir Pod Lateral Transfer SystemsSrs-ez mattress replacement cover with safety Side rail

ez turn banner

hospital product brochure download

Continuing Education Program

Continuing Education Units

For Clinical or Technical Assistance

 


Medical Professionals & Dealers
CALL 800 795-6115

For technical assistance or help choosing the correct product for your specific application.


virtual inservice avalable

 

Recent Blog

Blue Chip Medical Products celebrates its 27th year of wound care innovation.
Blue Chip Medical Products proudly commemorates its 27th anniversary as a pioneer in the field of wound prevention and treatment. With a steadfast commitment to excellence and patient care, Blue Chip continually innovates new products that consistently set the standard for clinically effective wound care across the America’s and Europe.  Through relentless research and development […]

Read More

The difference between Alternating Pressure and Air Fluidized Therapy AFT
    COMMON USE ALTERNATING PRESSURE Prevention & Treatment of Stages I-IV Pressure Injury AIR FLUIDIZED THERAPY Treatment of Stages III-IV Pressure Injury, burns, flap surgeries, grafts & unstageable wounds Both AFT and Alternating Pressure therapy are for the treatment of late-stage pressure sores. There are benefits and restrictions to each therapy. However, where, when, […]

Read More

What is Reactive Hyperemia and Why it's Important for Wound Care
Reactive hyperemia is the body’s response to periods of inadequate blood supply to a tissue or organ such as the skin. This phenomenon in wound care, refers to increase of blood circulation in the small arteries and capillaries after the skin has experienced a temporary obstruction in blood flow. Obstruction in blood supply to the […]

Read More

UPCOMING SHOWS

VA-MIAMI – SAFE PATIENT HANDLING & MOBILITY
MIAMI VA MEDICAL CENTER
MARCH 18 TH

MEDTRADE CONFERENCE
DALLAS TX
MARCH 26TH-28TH 2024

VA – SAFE PATIENT HANDLING
SALEM VA MEDICAL CENTER
MAY 16 TH

FIME
FLORIDA INTERNATIONAL MEDICAL EXPO
JUNE 19-21 ST

SER WOCN (WOUND CARE NURSING)
CHATTANOOGA, TN
AUGUST 22 ND – AUGUST 24 TH CHATTANOOGA

Recent Articles